DIGORA OPTIME SOFTWARE DOWNLOAD FREE
The stylish design of the USB data box incorporates the latest electronics onthe market, providing the user with the best possible image quality within a few seconds. There are two alternatives to be used as a Service terminal: See documentation of the test phantom for details. That can happen if the motor is running in wrong speed check that the short circuit wire in the motor cable is connected in correct pins. Bigger value moves image left in application window, smaller value right. It depends how many signal is still connected between K and user interface.
Uploader: | Zushakar |
Date Added: | 23 June 2018 |
File Size: | 50.54 Mb |
Operating Systems: | Windows NT/2000/XP/2003/2003/7/8/10 MacOS 10/X |
Downloads: | 47839 |
Price: | Free* [*Free Regsitration Required] |

Soredex is an imaging system also known as Digora. All updated files and related detailed instructions are described in the release letter of FW to be updated. Replacing of the bearings may pose unit to lowered image quality! Insert the new K board. Optional Import an image to show on the button 22 x 22 pixels. Under the copyright laws the documentation may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine readable form in whole or part, without the o;time written permission of SOREDEX.
Digital Imaging
Press the motor towards the wheel and push the belt off. For that move the plate carrier backwards and move the door by hand. The main cover must be attached on the backside by the screw. Detach the connectors, open the main cover and the light cover to check the door. Therefore, this document should not be regarded as an infallible guide to current product specifications. Then restart the unit and scan an image plate.
Smooth functions of the head support elements are comfortable for the patient.
Type Label Laser Warning Label Attach screw -if you open the main cover check that the label on the light cover is ok. Be very careful not to use service commands without the necessary know-how.

The belt is tightly on the wheel Tighten the belt using 3mm hex key. Belt tightness is now in lower opptime one ring is shown.

Do not use any other power supply units with the scanner. For more information or to arrange a free IT audit for your surgery contact Steve from Leading Dental on K is not a spare part.
Open Dental Software - Digora Soredex Optime Bridge
See Error Messages section. Check that there is no mechanical reason which prevents linear movement for example image plate or protective covers. Remove K board 6. Release K locking lever A in image and lift the board from K connector A 4.
DIGORA Optime DXR-60 Field Service Manual
If the opto is clean and the unit is still not working, replace K board. Switch softtware unit off and on to see that there are no errors during start up.
Icreasing exposure time does not really help. The imaging plates are flexible enough to enable easy and precise positioning xigora comfort for the patient, while being rigid enough to minimize distortion. Turn the unit on and start S2terminal Remove the lower shaft the same way as the upper.
DIGORA Optime Classic - Biotech Innovations Limited
SOREDEX reserves the right to make changes in specification and features shown herein, or discontinue the product described at any time without notice or obligation.
Check erase led current and voltage Color of sensitive surface changed. AutoEject AutoEject simplifies the workflow. After that test that start position is correct.